Certificate: CE, FDA, ISO 13485, RoHs, Reach
The infrared ear thermometer is specifically designed for safe use in the eardrum. The Infrared Ear Thermometer is a device capable of measuring people's body temperature by detecting the intensity of infrared light emitted from the ear canal of human. It converts the measured heat into a temperature reading displayed on the LCD. When properly used, it will quickly assess your temperature in an accurate manner.
S/N
|
Feature
|
Description
|
1
|
Product name
|
1 second reading digital medical baby forehead infrared thermometer
|
2
|
Model
|
ET-310
|
3
|
Size:
|
155x 38x 47mm
|
4
|
Main Features
|
10 reading memory, dual scale, cpmpact ergonomic size, beeps when ready, jumbo display, replaceable battery, auto-off, probe covers
|
5
|
Measuring site
|
Ear canal (Ear mode)
|
6
|
Measuring range
|
32.0℃~43.0℃ (89.6℉~109.4℉)
|
7
|
Reference body site:
|
Oral (This thermometer converts the ear temperature to display its “oral equivalent.”)
|
8
|
Operation mode:
|
Ear mode (Adjusted mode)
|
9
|
Laboratory accuracy
|
±0.2℃ (0.4℉) during 35.0℃~42.0℃ (95.0℉~107.6℉) at 15℃~35℃ (59.0℉~95.0℉) operating temperature range ±0.3℃ (0.5℉) for other measuring and operating temperature range
|
10
|
Display resolution
|
0.1℃ or 0.1℉
|
11
|
Measure time
|
Approximately two second
|
12
|
Operating temperature
range: |
10℃~40℃ (50℉~104℉), 15%~85%RH, non-condensing Atmospheric Pressure : 700hPa ~ 1060hPa
|
13
|
Storage and transport
temperature range |
-25℃~ 55℃ (-13℉~131℉), 15%~95%RH, non-condensing Atmospheric Pressure : 700hPa ~ 1060hPa
|
14
|
Warranty
|
2 year
|
15
|
Clinical accuracy
|
Clinical bias:0.12℃(0.2℉)
Clinical repeatability:0.12℃(0.2℉) Limits of agreement:0.8℃(1.4℉) |
16
|
Ingress protecting rating
|
IP22
|
Payment:T/T
Delivery way:EXW